There is a possible gap between CT and all PETS. You can check it thanks to the CT/PET Fusion displayed like the following :![]()
This gap is also computed and a warning is displayed like the following :
for IMAGERIE-C.H.U.-GRENOBLE_00_jaszczak_CT.nii mean shift (on three hottest spheres) between ct and pet : (2.9471324855151502, -2.2583466849401836, -0.11452005861740371) voxels
Note : If there is a gap, the camera should be fixed
display_fusion_PetCt: Choice ( input )
CT: Phantom CT ( input )
Pets_mean: ListOf( Phantom PET Mean ) ( input )
Toolbox : Nuclear Imaging
User level : 2
Identifier :
jzkCheckFusionPetCt
File name :
brainvisa/toolboxes/nuclearimaging/processes/jaszczak/jzkCheckFusionPetCt.py
Supported file formats :
CT :NIFTI-1 image, NIFTI-1 imagePets_mean :NIFTI-1 image, NIFTI-1 image